Past, present, and future of backscatter electron (BSE) imaging
Oliver C. Wells, Michael S. Gordon, et al.
ScMi 2012
Chemical-mechanical planarization (CMP) is a process that gives a flat surface on a silicon wafer by removing material from above a chosen level. This flat surface must then be reviewed (typically using a laser) and inspected for scratches and other topographic defects. This inspection has been done using both the atomic force microscope (AFM) and the scanning electron microscope (SEM), each of which has its own advantages and disadvantages. In this study, the low-loss electron (LLE) method in the SEM was applied to CMP samples at close to a right angle to the beam. The LLEs show shallower topographic defects more clearly than it is possible with the secondary electron (SE) imaging method. These images were then calibrated and compared with those obtained using the AFM, showing the value of both methods. It is believed that the next step is to examine such samples at a right angle to the beam in the SEM using the magnetically filtered LLE imaging method.
Oliver C. Wells, Michael S. Gordon, et al.
ScMi 2012
Oliver C. Wells
Review of Scientific Instruments
Oliver C. Wells
Journal of Microscopy
Oliver C. Wells
Scanning